2 Press Height. Check the central positioning of the uptake needle by using the
buttons Move up and Move down.
3 Select Position and adjust by using the Move back and Move fwd buttons.
Select Done after making necessary adjustments.
Figure 5.10: Calibrate the uptake needle position in the washing station.
4 Select Save to save the new configuration.
5.1.1.3 Calibrate the washing station: NEG needle (step 2/5)
1 Select Calibrate to proceed with calibration.
Figure 5.11: Calibration 1 Step 2/5: check NEG needle position in the washing station.
2 Chose either current position to start with (recommended, press Use) or
reset needle position to factory settings (press Reset). The needle arm will
automatically move towards the washing station. The needle should be located
directly above the center of the front opening of the washing station (see
figure 5.5).