6.11 Shutting down the instrument
If the instrument is to be shut down after separation, the Sleep program can be
chosen as a wash program after the last separation instead of Rinse or Qrinse.
However, if Sleep is chosen as a wash program, the autoMACS Pro Separator will
not allow definition of any programs beyond this position. Alternatively, press the
shutdown button on the upper right hand corner of the screen and confirm. Upon
completion of the Sleep program, the fluid bottles are illuminated purple. Switch
off the autoMACS Pro Separator using the main power switch. Note: The autoMACS
Pro Separator automatically performs the Sleep program if the device is inactive for
more than six hours. For long-term storage of the instrument, refer to section 3.2.2.