aerosol cleaning agents; always use a damp cloth.
The instrument is equipped with a threewire electrical groundingtype plug that
has a third pin for grounding. This plug only fits into a grounded power outlet. This is
a safety feature. Do not try to insert the plug into a nongrounded power outlet. If
you cannot insert the plug into the outlet, contact your local electrician to replace
the outlet.
The instrument should only be operated from a power source indicated on the
product’s electrical ratings label. If you have questions about the type of power
source to use, contact your authorized Miltenyi Biotec service provider or local
power company. Do not use extension cords or power strips. Do not overload an
electrical outlet. The overall system load must not exceed 80% of the branch circuit
rating.
Make sure that the main switch as well as the connector for the power cable are
easily accessible and located as close to the operator of the instrument as possible.
If it is necessary to disconnect the power supply, unplug the cable from the power
outlet.
Only peripheral devices that comply to UL 60950 are allowed to be connected to the
RS232 connector labeled COM. The connector labeled RS232/AUX is not in use. In
addition, only original autoMACS Pro Separator equipment should be attached to
the connectors labeled External CAN, CAN1, and CAN2. The voltage levels on these
connectors shall not exceed hazardous voltage levels of 30 V RMS. and 42.4 V peak
or 60 Vdc. Only the autoMACS Pro Bottle Sensor Cable should be attached to the
Bottle Sensor connector. Only a 2D code reader recommended by Miltenyi Biotec
should be connected to the RS232/BCR connector. External laser devices connected
to the connector labeled RS232/BCR have to comply with the standard IEC 608251.
Only use connector cables less than 3m in length.
Strong magnetic field
The autoMACS Pro Separator is equipped with an extremely powerful magnet.
There is a risk of severe personal injury for persons carrying pacemakers,
brain shunts, or electronic medical implants. Keep any magnetic information
carriers (such as credit cards, magnetic tapes, and floppy disks), any electronic
equipment (such as hearing aids, pacemakers, measuring and control instruments,
computers, and watches), and magnetizable tools and objects at a distance of at
least 20 cm from the magnet cover. These items may be affected or damaged by the
magnetic field.