155
TECHNIC
AL
D
AT
A
AND
SPECIFIC
ATIONS
10
10.1 Technical data and specifications
WARNING! Read the chapter Important safety information before installation
and use.
The autoMACS Pro Separator is labeled as a protection class I instrument and must
be plugged into a grounded power outlet. The MACS MiniSampler is labeled as a
protection class III instrument and must only be plugged into the connector
labeled with External CAN of the autoMACS Pro Separator.
The main power supply cord and plug of the instrument shall comply with
following specifications (USA and Canada only): UL listed and KAM cord, minimum
type SJ, minimum 18 AWG, 3 conductors. Rated for a minimum temperature of
60 °C. Provided with grounding-type (NEMA 5-15P) attachment plug, rated 125 Vac,
10 A. Opposite end terminates in IEC 320 style connector, rated 125 Vac, 10 A.
Conditions of operation: 15–30 °C with 0–85% humidity (non-condensing) at a
maximum altitude of 2000 m. Supply voltage fluctuations up to +/-10% of the
nominal voltage. Transient overvoltages present on the mains supply: category II.
The instrument is suitable for rated pollution degree 2. The autoMACS Pro
Separator is not specified for use in the cold room.
The emission sound pressure level at the workstation is <70 dB(A).