4.5.2 User settings
User settings contain programs necessary for maintenance and setup of the
autoMACS Pro Separator.
About – System and hardware information
This program informs the user about the software version, serial number of the
instrument, and other hardware information.
Calibr_1 – Calibrating the needle arm position
This program is used for the calibration of the needle arm to the washing station
and tube racks. Refer to section 5 for more information.
Calibr_2 – Calibrating the liquid volume control
This program is used for the calibration of the liquid volume control. Refer to
section 5 for more information.
Check_up – Performing a system check-up
The Check_up program allows the user to perform a system check-up. It is
recommended to use the program if hardware errors occur. Refer to section 9.1
for details.
O_bcr – 2D code reader setup, configuration, and initialization
The O_bcr program allows the user to setup, configure, and initialize the autoMACS
Pro Separator 2D code reader.
O_init – Optional priming of the instrument at startup
By default, the autoMACS Pro Separator does not perform a wash program after
initialization. The option O_init allows the user to add an initial rinse program to
automatically prime the instrument after initialization.
O_led – Activate/deactivate fluid bottle illumination
The autoMACS Pro Separator has a bottle illumination designed to facilitate
monitoring the instrument’s status. The bottle illumination can be switched
on or off.