37091/04 – ch20 (Issued: 2020-06)
CliniMACS® Plus System User Manual (Canadian edition)
4 – STEP
3: CliniMACS® Tubing Set and CliniMACS Tubing Set LS
Insert separation column and load valve no. 5
The window will display Screen 4-3.2 as shown.
The valves shown on the screen will be opened automatically.
1. Insert the separation column into the separation column
holder as shown (see Figure 4-3.3).
Note
To avoid possible pinch injury, insert the separation column
as follows: Hold the top and bottom of the column between
thumb and index finger, then carefully insert the separation
column into the separation column holder.
2. Load the tubing into valve no. 5.
To proceed, press
ENT
Note
■
As each step is performed, check all tubing and attachments
for any kinks or severe bending that could restrict the flow
of liquid through the tubing. Check all valves to ensure the
tubing fits snugly.
■
Only insert the tubing set into open valves (when button is
pushed inwards). The tubing will not fit correctly if inserted
into a closed valve.
■
If the tubing has to be adjusted after a valve has been closed,
do not pull the tubing without pressing the valve button to
open the valve (see Figure 4-3.4).
Screen 4-3.2: Insert separation column and load
valve no. 5
Figure 4-3.3: Separation column in separation
column holder
Figure 4-3.4: Correctly inserted tubing
11
Press button
to open valve.
Tubing
Inner end
of slot