STEP 3
Installation of the CliniMACS®
Depletion Tubing Set
Follow the installation instructions given on the screens of
the CliniMACS® Plus Instrument.
(1) Install upper part.
(2) Install lower part.
Proceed to the priming procedure.
(3) Perform integrity test.
WARNING:
■
To perform the integrity test, clamp the tubing
below the Non-Target Cell Bag, before starting the
integrity test.
■
Remove the clamp after completion of the integrity
test.
(4) Connect the Cell Preparation Bag:
– Firmly insert the spike of the pre-system filter into
the tubing set,
– Connect pre-system filter and spike connector,
– Spike Cell Preparation Bag,
– Ensure that the connections are secure and that
free liquid flow is possible.
(5) Level bags:
VERY IMPORTANT
Ensure that the bu
ff
er bag, the Cell Preparation Bag,
the Non-Target Cell Bag, and the Reapplication Bag are
leveled correctly as displayed and that the bags and
tubings are neither stretched nor bent.
Highest position:
Bu
ff
er bag
Middle Position:
Reapplication Bag and
Non-Target Cell Bag
Lowest position:
Cell Preparation Bag
1
2
4
5
3
Installation of the CliniMA
CS® D
epletion T
ubing S
et
37091/04
ch20
(Issued:
2020-06)