Connect and switch on the instrument
Connect the instrument to an uninterruptible power supply
using the supplied power cord. Switch on the instrument by
using the main ON/OFF switch (see Figure 1.5) located on the
right hand back panel of the instrument.
For safety, the instrument should be turned o
ff
after each run
and the power cord cable should be disconnected during the
instrument clean-up procedures.
Upon start up, the program will automatically be loaded and
Screen 1.1 will appear in the display window.
If the instrument does not start up, switch the instrument o
ff
and disconnect it from the power supply. Check the power cord
connection and the fuses, which are located on the right hand
back panel. Then switch on the instrument again.
If the instrument does not start correctly or the window displays
an error message, note the error message number, switch
the instrument o
ff
and contact the Miltenyi Biotec Technical
Support.
Proper training is required to operate the instrument. Read
the instructions carefully. Further training by Miltenyi Biotec
authorized representative may be required. This instruction
includes details on instrument operation, sample handling, and
troubleshooting.
Language selection and service menu
The instument provides a menu to set-up the language and a
service menu.
To enter these menus wait until Screen 1.1 appears in the display
window and DO NOT press ‘ENT’ as shown on the window
display.
To start the general menus, refer to section 1.3.8.
Screen 1.1: Main screen
37091/04 – ch20 (Issued: 2020-06)
CliniMACS® Plus System User Manual (Canadian edition)
1 – 16