Parameter
Specification
Minimum sample volume
1 μL (recommended)
Minimum sample particle size
1 μm
Sample flow rate
25-100 μl/min plus automated flow rate to maintain 500, 1000,
or 2000 events/second
Sheath fluid consumption
2-10 mL/min
Maximal event rate
Up to 15,000 events/second
Lasers MACSQuant Analyzer 10
488 nm, 404 nm, 640 nm
Lasers MACSQuant VYB
561 nm, 404 nm, 488 nm
Lasers MACSQuant Analyzer 16
404 nm, 488 nm, 642 nm
Fluorescence sensitivity
MESFs (CV <5%): FITC < 200; PE < 100; APC < 150
Absolute cell count performance
Ac/- 7%; reproducibility (CV) <5%
Sample carryover
Typical value of 0.01%
Table 9.2:
Operation details of the MACSQuant Analyzer 10, VYB and Analyzer 16.