3.5.3
Collecting the unlabeled fraction
1
For application of samples and buffer the MultiMACS Cell24 Magnet and the MultiMACS Column Holder
move down.
2
Apply buffer or sample depending on the process step.
Equilibrate columns just before the separation. Do not store equilibrated columns.
3.5.4
Tip-Touch
The Tip-Touch Plate is used for the removal of residual drops from the column tips. The plate can be moved
back and forth to deposit the last drop into the wells. This ensures reproducible recoveries and avoids cross-
contamination. The column tips must protrude 1 mm into the well.
1
MultiMACS Column Holder and MultiMACS Cell24 Magnet remain down for the Tip-Touch process.
2
When the instrument prompts you to perform a Tip-Touch, move the Tip-Touch Plate firmly back and forth
once. Droplets from the column tips fall into the plate.
Tip-Touch can be ignored if single-well deep well plates are used.
3
O
P
E
R
A
T
IN
G
T
H
E
IN
S
T
R
U
M
E
N
T
52
Summary of Contents for MultiMACS Cell24 Separator Plus
Page 1: ...MultiMACS Cell24 Separator Plus User manual...
Page 8: ...6...
Page 16: ...IMPORTANT SAFETY INFORMATION EN DE ES FR 14...
Page 24: ...WICHTIGE SICHERHEITSINFORMATIONEN EN DE ES FR 22...
Page 32: ...INFORMACI N IMPORTANTE DE SEGURIDAD EN DE ES FR 30...
Page 40: ...INFORMATIONS DE S CURIT IMPORTANTES EN DE ES FR 38...
Page 44: ...1 INTRODUCTION 42...
Page 66: ...5 PRE DEFINED PROCESSES 64...
Page 72: ...6 CUSTOM PROCESSES 70...
Page 78: ...7 MAINTENANCE 76...
Page 86: ...9 TECHNICAL DATA AND SPECIFICATIONS 84...
Page 88: ...10 TECHNICAL SUPPORT 86...
Page 91: ......