7
M
A
IN
T
E
N
A
N
C
E
75
3
Scroll down to
LOCK UNIT FOR TRANSPORT
.
4
Tap
LOCK UNIT FOR TRANSPORT
. The MultiMACS Cell24 Magnet and MultiMACS Column Holder will
move to the lock position.
5
When
MULTIMACS SEP. IS READY FOR TRANSPORT. SWITCH DEVICE OFF
is shown, switch off the
instrument.
6
Detach the Tip-Touch Plate.
7
Disconnect the MACS Elution Chamber. Push the tube connection ring towards the MACS Elution Chamber
and simultaneously pull out the tube connector.
8
The MultiMACS Cell24 Separator Plus is now ready for shipping.