Troubleshooting
8.1
Cross contamination by drops at the column tips
Always perform a Tip-Touch before removing the columns from the MultiMACS Column Holder. The columns
must move slightly by the contact.
1
Ensure that all drops at the column tips have been removed after the Tip-Touch. If drops are still present, tap
MOVE BACK
and repeat the Tip-Touch.
2
Tap
OK
.
Clean the feet of the MultiMACS Cell24 Separator Plus if the instrument moves during Tip-Touch (refer to
Cleaning the instrument on page 71
8.2
Collision detection
If a tilting of the MultiMACS Cell24 Separator Plus is registered while the MultiMACS Cell24 Magnet is moving,
the movement stops and the following screen is shown.
1
If no blocking objects are visible, tap
RETRY
.
2
If a blocking object is visible, tap
MOVE BACK
to set back the MultiMACS Cell24 Magnet to its former
position.
3
Take out any objects causing the stop of the movement.
4
Tap
OK
to restart the move.
8.3
Vacuum failure
During generation of the vacuum the status LED of the MACS Elution Station blinks green or red. When the
vacuum is established, the blinking stops and the LED lights up constantly green.
A constant red status LED of the MACS Elution Station indicates a failure of function. Contact Miltenyi Biotec
Technical Support if red blinking of the LED persists or the LED lights up constantly red.
Summary of Contents for MultiMACS Cell24 Separator Plus
Page 1: ...MultiMACS Cell24 Separator Plus User manual...
Page 8: ...6...
Page 16: ...IMPORTANT SAFETY INFORMATION EN DE ES FR 14...
Page 24: ...WICHTIGE SICHERHEITSINFORMATIONEN EN DE ES FR 22...
Page 32: ...INFORMACI N IMPORTANTE DE SEGURIDAD EN DE ES FR 30...
Page 40: ...INFORMATIONS DE S CURIT IMPORTANTES EN DE ES FR 38...
Page 44: ...1 INTRODUCTION 42...
Page 66: ...5 PRE DEFINED PROCESSES 64...
Page 72: ...6 CUSTOM PROCESSES 70...
Page 78: ...7 MAINTENANCE 76...
Page 86: ...9 TECHNICAL DATA AND SPECIFICATIONS 84...
Page 88: ...10 TECHNICAL SUPPORT 86...
Page 91: ......