133
8 Troubleshooting
Error code table
All errors are coded. Refer to the following table for explanations.
Displayed
error message
Explanation and remedy
STEPPING LOSS
ERROR –101
INITIALIZING
SYSTEM
Motor out of sequence. Touch OK for
intermediate reinitialization of the system.
WRONG MAG POS
ERROR –102
INITIALIZING
SYSTEM
Incorrect positioning of the magnet. Touch OK
for intermediate reinitialization of the system.
WRONG COL POS
ERROR –106
INITIALIZING
SYSTEM
Column Holder is locked in wrong position.
Touch OK for intermediate reinitialization of the
system. Column holder will be repositioned.
POSITION FAILED
ERROR –111
INITIALIZING
SYSTEM
Incorrect positioning of the magnet. Make sure
device is not exposed to direct sunlight. Touch
OK for intermediate reinitialization of the
system.
CANNOT PICKUP
ERROR –117
INITIALIZING
SYSTEM
Locking of Column Holder did not work.
Touch OK for intermediate reinitialization
of the system.
WRONG SIGNAL
ERROR –118
CALL TECHNICAL
SERVICE
Electronic hardware problem.
Call technical service.
LOCK FAILURE
ERROR –123
CALL TECHNICAL
SERVICE
Electronic or mechanical hardware problem.
Call technical service.
MAGN. SENSOR
ERROR –205
Electronic hardware problem.
Call technical service.
Table 8.2: Explanation of error code and suggestion for fixing.
Summary of Contents for MultiMACS M96
Page 1: ...Original instructions MultiMACS M96 Separator User manual...
Page 28: ...27...
Page 40: ...39...
Page 52: ...51...
Page 64: ...63...
Page 68: ...67...
Page 72: ...71...
Page 76: ...75...
Page 80: ...79...
Page 127: ...126 6 Instructions for use Figure 5a Landscape orientation Figure 5b Portrait orientation...
Page 128: ...127...
Page 130: ...129...
Page 136: ...135...
Page 138: ...137...
Page 142: ......
Page 143: ......