134
8 Troubleshooting
8
Fuse holder
Power cable
connector
Fuse exchange
First disconnect both the power and the communication cable.
The fuse holder is located on the rear side of MultiMACS M96
Separator at the power connector, refer to figure 6. Detach the
fuse holder with a sharp tool, e.g., a screwdriver. Remove the
defect fuse and insert a new fuse with the type and rating
5×20 mm, T2.5AH250V. Reinsert the fuse holder.
Caution:
Replace fuses only with same type and rating to ensure
protection against fire risk.
Figure 6: Fuse holder on the rear side of the MultiMACS M96 Separator.