3 Technical data and specifications
Technical data and specifications
Technical data
Technical data
Specification
Model
MultiMACS M96 Separator
Order number
130-091-937
Dimension (w × d × h)
230 mm × 435 mm × 230 mm
Weight
11 kg
Input voltage
100 – 240 V ~, 50/60 Hz
Power consumption
200 W
Electric current
3.5 A max.
Fuses
T3.15AH250V, 5 × 20 mm
RS232-Interface (labeled “com”)
Pin 1, 4, 6, 7, 8, 9 NC
Pin 2 RXD
Pin 3 TXD
Pin 5 GND
DC-Output
Module 1 24 VDC / 8 A
Module 2 24 VDC / 0.315 A
Working temperature
15–30 °C
Humidity
0–85%
Altitude
max. 2000 m
Emission sound pressure level at
workstation
<70 dB(A)
Table 3.1: Technical data of the MultiMACS M96 Separator.
The MultiMACS M96 Separator is labeled as a protection class I device
and must be plugged into a grounded power outlet, refer to chapter 1,
General precautions.
The MultiMACS M96 Separator is not specified for use in the cold
room. The MultiMACS M96 Separator can process up to 96 samples
in parallel. The MultiMACS M96 Separator has been investigated in
accordance with the standards UL 61010-1 and CAN/CSA C22.2
No. 61010.1.
Compliance was demonstrated by conformance to the following
specification FCC 47 CFR Ch. 1 part 15, class B.
Summary of Contents for MultiMACS M96
Page 1: ...Original instructions MultiMACS M96 Separator User manual...
Page 28: ...27...
Page 40: ...39...
Page 52: ...51...
Page 64: ...63...
Page 68: ...67...
Page 72: ...71...
Page 76: ...75...
Page 80: ...79...
Page 127: ...126 6 Instructions for use Figure 5a Landscape orientation Figure 5b Portrait orientation...
Page 128: ...127...
Page 130: ...129...
Page 136: ...135...
Page 138: ...137...
Page 142: ......
Page 143: ......