6 Instructions for use
5. Insert the waste plate, e.g., the Deep Well Block, 2.5 mL, included
in the kit, onto the Tip-Touch Plate. If using a plate with a different
height adjust the height of the waste plate as described in chapter 6,
Editing a process.
Touch
OK
and the MultiMACS 96 Magnet will move downwards until
column tips slightly immerse in the waste plate. The next screen
appears.
RINSE, APPLY
SAMPLE, WASH.
IF REQUIRED:
PRE-ELUTE
MOVE
BACK
OK
ESC
6. Rinse columns with equilibration buffer, apply sample, and wash.
A pre-elution might be required to exchange the void volume of the
column, refer to the respective kit data sheet for detailed
instructions. Touch
OK
for the following screen.
TIP-TOUCH
COLUMNS IN
PLATE
OK
ESC
7. Move the Tip-Touch Plate firmly back and forth once so that the
inner walls of the wells of the Deep Well Block touch the tips of the
Multi-8 Columns; thereby, any drop on the column tips that did not
fall by gravity is removed.
Touch
OK
and the MultiMACS 96 Magnet will move upwards to allow
removal of the waste plate.
INSERT
ELUTION PLATE:
MTP (14 MM)
MOVE
BACK
OK
ESC
Summary of Contents for MultiMACS M96
Page 1: ...Original instructions MultiMACS M96 Separator User manual...
Page 28: ...27...
Page 40: ...39...
Page 52: ...51...
Page 64: ...63...
Page 68: ...67...
Page 72: ...71...
Page 76: ...75...
Page 80: ...79...
Page 127: ...126 6 Instructions for use Figure 5a Landscape orientation Figure 5b Portrait orientation...
Page 128: ...127...
Page 130: ...129...
Page 136: ...135...
Page 138: ...137...
Page 142: ......
Page 143: ......