6 Instructions for use
96 POS + BEAD
Caution: Read chapter 4, Safety functions, carefully before running a
process! Then follow the instructions in chapter 4, Semi-automated
bench-top use. Always be sure to use MultiMACS 96 Magnet,
MultiMACS
Column Holder, and the plates in the same orientation as
described in chapter 6, Rotating MultiMACS 96 Magnet and
Column Holder.
The MultiMACS M96 Separator contains a list of pre-defined
separation programs to choose from, visit www.miltenyibiotec.com
for updates. 96 POS + BEAD is used for applications where the
target fraction is the molecule-MicroBead complex. The target
is eluted outside the magnet.
If a process with different process parameters has to be run, either
create a new process, as described in chapter 6, Creating a new
user-defined process, or, if only a few parameters have to be
changed, edit the parameters as described in chapter 6, Editing a
process. For further details on user-defined processes, refer to
chapter 6, Running a user-defined process.
1. Switch on the instrument and touch the Welcome Screen or
wait for a few seconds for the Process Selection Screen.
WELCOME TO THE
MULTIMACS SEPARATOR
BY MILTENYI BIOTEC
96 POS + BEAD NEW
96 NEG/POS
SET
UP
The last process performed on the MultiMACS M96 Separator is
displayed on the upper left (default: 96 POS). The one but last
process is listed below.
If necessary scroll through the list of available process names by
touching the symbol or until 96 POS+BEAD is displayed.
Summary of Contents for MultiMACS M96
Page 1: ...Original instructions MultiMACS M96 Separator User manual...
Page 28: ...27...
Page 40: ...39...
Page 52: ...51...
Page 64: ...63...
Page 68: ...67...
Page 72: ...71...
Page 76: ...75...
Page 80: ...79...
Page 127: ...126 6 Instructions for use Figure 5a Landscape orientation Figure 5b Portrait orientation...
Page 128: ...127...
Page 130: ...129...
Page 136: ...135...
Page 138: ...137...
Page 142: ......
Page 143: ......