Adjusting the horizontal focus
1
To adjust the horizontal focus, select the Calibration tab on the bottom of the Settings 1 window.
2
Move the slider Light Sheet Focus > Offset Calibration until only one beam passes through the pinhole.
Ensure that the horizontal focus is always in the center and the Light Sheet Properties > Sheet NA is at the
maximum. If necessary, readjust the z-direction.
3
If a clearly recognizable, single beam passes through the pinhole in the center of the mask, increase the
intensity of this beam by moving the alignment tool in z-direction. The central beam is now adjusted to the
highest intensity. Repeat this procedure until no more improvement is possible.
4
Click Save Settings in the toolbar to fix the adjusted horizontal focus.
After the settings have been saved, the alignment tool must not be moved again. It is highly recommended
to block the third button from the left of the jog wheel.