3.6
Unpacking and installation
Protect the instrument from shock and vibration. Do not place vibrating devices, such as vortex mixers, on
the same table as the instrument. This could impair the quality of the results.
Only Miltenyi Biotec service personnel is allowed to unpack and install the instrument. Contact Miltenyi Biotec
Technical Support for further information.
Before installation, ensure that your site is properly prepared. The instrument should be installed on an optical
table with vibration isolation that can carry a weight of at least 250 kg. Allow a minimum of 20 cm free space to
the rear side and to both sides of the instrument. Allow a minimum of 100 cm free space in front of the
instrument. All sides of the instrument must be accessible for service and maintenance. Due to vibrations, an
adjacent desk is recommended for the laser source, the workstation PC and its components.
3
IN
ST
A
LL
A
TI
O
N
22