6.2
Placing the alignment tool
1
Attach the alignment tool with the sample holder to the sample holder frame.
2
Place the sample holder frame into the cuvette.
6.3
Setting the light sheet path
LASER RADIATION
Direct exposure to laser beam may lead to eye injury.
• Do not stare into laser beam.
• Do not insert reflective objects into the cuvette.
• Do not manipulate the alignment tool during alignment.
Before setting the light sheet path, ensure that not any bubbles are present to the sides of the alignment tool.
1
Mount the appropriate dipping cap.
2
Raise the sheet width temporarily to locate the light sheets.
3
Click the
Start live preview
button in the
Measurement Wizard
window to start the live preview. The
button turns into a yellow square.
6
LA
SE
R
A
LI
G
N
M
EN
T
30