MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT 

 

REPAIR SERVICE INSTRUCTIONS — CALIBRATION 

2266-20 Laser TEMP-GUN™ Thermometer 

 

Environmental 

Condition 

Perform all calibration at an ambient temperature of 23°C ± 2°C / 73.4°F ± 3.6°F and relative 

humidity  of  80%  -  allow  the  TEMP-GUN™  thermometer  to  sit  at  this  temperature  for  a 

minimum of 30 minutes before proceeding. 

 

 
 

REPAIR SERVICE 

INSTRUCTIONS 

Calibration 

 

 

 

 

 

 

 

 

 

 

 

 

 

MILWAUKEE ELECTRIC TOOL CORPORATION 

13135 WEST LISBON ROAD 

 BROOKFIELD, WISCONSIN 53005-2550 

 USA 

58-92-2266d1   04/2010

Summary of Contents for 2266-20 Laser TEMP-GUN

Page 1: ...erform all calibration at an ambient temperature of 23 C 2 C 73 4 F 3 6 F and relative humidity of 80 allow the TEMP GUN thermometer to sit at this temperature for a minimum of 30 minutes before proce...

Page 2: ...Specifications 3 Basic Maintenance 4 Loading Changing the Batteries 4 Maintaining Tool 4 Cleaning the Laser Windows 4 Scope 5 Equipment Use 5 Testing Condition and Interference 5 Field calibration Pr...

Page 3: ...nual If you do not do so the protection provided by the Laser TEMP Gun may be impaired Read the Safety Information page before servicing this product In this manual a WARNING identifies conditions and...

Page 4: ...ctices that could result in personal injury or loss of life Specific precautions To reduce the risk of injury user must read and understand operator s manual Laser light Do not stare into beam or view...

Page 5: ...2 C 4 F or 2 of reading whichever is greater Assume ambient operating temperature of 23 C to 25 C 73 F to 77 F Min measuring distance 2 50 C 122 F 4 50 C 122 F Display resolution 0 1 C 0 1 F Contact t...

Page 6: ...r operate at full power with fully charged batteries clean the contacts on the battery door If the tool still does not work properly return the tool to a MILWAUKEE service facility for repairs Cleanin...

Page 7: ...40 1 10 in a controlled room Each thermometer should be uniquely identified as to the room andlor area in which it is used or tested 2 2 If the meter is exposed to significant changes in ambient tempe...

Page 8: ...mometer as shown in Figure 1 if necessary 3 1 6 Position correctly the IR opening towards the black body target source at 3 inches apart for Fluke 9133 or 11 inches for Fluke4181 Ensure that the senso...

Page 9: ...ration mode 3 2 4 Press ALM button to skip 25 C as this is for factory calibration only 3 2 5 32 F or O C degrees will then appear for next calibration point 3 2 6 Set the Fluke 714 724 calibrator as...