16 Probes and Biopsy
Operator’s Manual
16 - 33
NGB-018 Metal-needle detachable needle-guided bracket
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Select a suitable needle-guided bracket and match the groove to the tab of the probe.
Mount the bracket onto the probe.
c.
Screw the pinch nut of the needle-guided bracket to ensure that the needle-guided bracket
is properly installed on the probe.
2.
Adjust the needle angle to the proper shift as required:
a.
Loosen the angle pinch nut.
b.
Adjust the angle block to the desired level.
c.
Tighten the angle pinch nut.
1
Pinch nut of bracket
2
Angle pinch nut
3
Angle block
4
Clamp
5
Groove
6
Angle-adjusting base
7
Angle shift sign
8
Guiding block specification
9
Guiding block
10
Needle guide hole
1
2
3
4
5
6
7
8
9
10
Summary of Contents for Ana
Page 2: ......
Page 50: ...This page intentionally left blank...
Page 60: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 116: ...This page intentionally left blank...
Page 166: ...This page intentionally left blank...
Page 176: ...This page intentionally left blank...
Page 194: ...This page intentionally left blank...
Page 220: ...This page intentionally left blank...
Page 288: ...This page intentionally left blank...
Page 304: ...This page intentionally left blank...
Page 308: ...This page intentionally left blank...
Page 316: ...This page intentionally left blank...
Page 337: ......
Page 338: ...P N 046 018835 00 2 0...