16 Probes and Biopsy
Operator’s Manual
16 - 41
NGB-043 Metal-needle detachable needle-guided bracket
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold the probe in one hand, select the correct needle-guided bracket and hold it with the
other hand. Match the grooves of the needle-guided bracket to the tabs of the probe.
c.
Mount the bracket onto the probe, and rotate the locking nut to fix the bracket.
2.
Insert the biopsy needle:
a.
Insert a biopsy needle into the guiding block hole.
b.
Adjust the biopsy angle of the needle, and ensure that the needle can be adjusted
continuously within the range of 30º to 85º.
3.
Release the needle from the bracket:
Hold the probe, and remove the biopsy needle form the guiding hole of the guiding block.
4.
Remove the needle-guided bracket:
1
Locking nut of left and right clamps
2
Groove
3
Guide frame
4
Guide block hole
5
Clamps
1
2
3
4
5
Summary of Contents for Ana
Page 2: ......
Page 50: ...This page intentionally left blank...
Page 60: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 116: ...This page intentionally left blank...
Page 166: ...This page intentionally left blank...
Page 176: ...This page intentionally left blank...
Page 194: ...This page intentionally left blank...
Page 220: ...This page intentionally left blank...
Page 288: ...This page intentionally left blank...
Page 304: ...This page intentionally left blank...
Page 308: ...This page intentionally left blank...
Page 316: ...This page intentionally left blank...
Page 337: ......
Page 338: ...P N 046 018835 00 2 0...