16 Probes and Biopsy
Operator’s Manual
16 - 49
16.2.4 Starting the biopsy procedure
DANGER
•
Ensure that all guide parts are properly fixed prior to performing a biopsy.
•
If you changed the probe or needle-guided bracket during the biopsy, verify
the guide line again.
•
Failure to match the guide zone displayed to the guide may cause the
needle to track a path outside the zone.
•
It is extremely important that when using the adjustable angle biopsy
guides, the angle displayed on the screen matches the angle set on the
guide, otherwise the needle will not follow the displayed guide zone and
this could result in repeated biopsies or patient injury.
Perform the following procedure:
1.
Select the correct needle-guided bracket and needle and install them properly.
2.
Tap [Biopsy] to enter the biopsy.
If the current probe has no corresponding bracket, or the image is frozen and the guide line was
hidden before the image was frozen, then you cannot enter the Biopsy menu.
3.
Select the bracket and guide line according to the actual situation.
Tap [Verify] to enter Biopsy Verify menu to fine tune the guide line if needed.
4.
If available, use iNeedle function to help enhance the needle visualization when the needle
display is not clear.
iNeedle is an option.
a.
Tap [iNeedle] in B menu.
The parameters that can be adjusted appear on the menu.
b.
Tap [B/iNeedle] to display B image and iNeedle image synchronously.
c.
Tap [iNeedle] again to exit iNeedle.
5.
Scan to locate the target. Center the target in the electronic guidezone path.
6.
Direct the needle into the area of interest for specimen.
7.
After extracting the biopsy sample is complete, gently remove the probe from the body.
8.
Tap [Exit] to exit the Biopsy menu
9.
Disassemble the items and properly dispose of these items as required.
16.2.5 Clean and Sterilize the Needle-Guided Bracket
CAUTION
•
Needle-guided brackets whose name starts with NGB are reusable, and
need thorough cleaning and sterilization before and after each biopsy.
•
Follow local regulations when selecting and using the disinfectant.
•
Repeated sterilization may degrade the safety and performance of the
needle-guided bracket. Before use, please check whether the needle-
guided bracket has defects such as deformation and rusting. If such
Summary of Contents for Ana
Page 2: ......
Page 50: ...This page intentionally left blank...
Page 60: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 116: ...This page intentionally left blank...
Page 166: ...This page intentionally left blank...
Page 176: ...This page intentionally left blank...
Page 194: ...This page intentionally left blank...
Page 220: ...This page intentionally left blank...
Page 288: ...This page intentionally left blank...
Page 304: ...This page intentionally left blank...
Page 308: ...This page intentionally left blank...
Page 316: ...This page intentionally left blank...
Page 337: ......
Page 338: ...P N 046 018835 00 2 0...