6 Image Acquisition
Operator’s Manual
6 - 15
Figure 6-1 Quantitative analysis display (taking velocity-time curve as the example)
Perform the following procedure:
1.
Scan the image with the moves of myocardium on, freeze the image and select the scan scope,
or open the image which includes the myocardium moves already.
1
TDI review
Sampling area: indicates the sampling position of the curve. The
sampling lines are marked with color numbers. It can mark 8
ROIs at most.
2
2D grey image review
• Use the trackball/trackpad; the images in TDI review
window and 2D review window are reviewed synchronously,
since the two images are frozen at the same time.
• ROI movement is linked between the TDI (Tissue Doppler
Imaging) review window and the 2D imaging reviewing
window.
3
Display analysis curve
• Y-axis represents the velocity (unit: cm/s) [in strain-time
curve, Y-axis represents the strain (%); in strain-time curve,
Y-axis represents the strain (unit: 1/s)].
• X-axis represents time (s);
• Frame mark: a white straight line perpendicular to the X-
axis, and can be moved left and right by using the trackball/
trackpad.
• Click the check box in front of the ROI to display or hide the
analysis curve.
• You can get the current X/Y axis value by moving the cursor
onto one point on the curve; and if you press <Set> at this
time, the frame marker will move to the spot.
4
ECG display area
/
2
1
3
4
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...