6 Image Acquisition
Operator’s Manual
6 - 33
–
Circumferential Rotation Rate curve: The X-axis represents time (s); the Y-axis represents
rotation by time (Deg/s).
Torsion & Torsion Rate Curve
The system provides left ventricular torsion data based on short axis sections of PSAX AP and
PSAX B. Torsion is acquired by calculating difference of apex and base of the heart.
Torsion = PSAX AP Rot. - PSA XB Rot.
•
The X-axis represents time (s).
•
The Y-axis represents tortion by time (Deg/s).
6.15.6 Bulleye
After tracking, the system can display Bull’s Eye graph, so as to judge reverse movement or scope
of myocardium.
1.
Tap [Bull’s Eye] on the touch screen to turn on the function:
You can acquire:
–
Time to peak value and peak value of the 17 segments (similar to 16 segments);
–
Display measurement result EDV/ESV/EF/TPSD.
2.
Use [Parameter] on the touch screen to see different parameter bull’s eye graph.
“-” will display in the table to indicate those segments that are not well tracked.
6.15.7 Measurement/Comment
Under tissue tracking QA mode, only Time measurement is available. For details, please refer to
Operator’s Manual
Advanced Volume
.
Comments and Body Mark operations are the same as in other modes.
6.15.8 Data Export
The system provides data exporting function, so that you can export calculation result for analysis
(for instance, SPSS analysis).
Tap [Data Export] on the touch screen, to export analyzed data of each segment in .csv format.
6.16 iWorks (Auto Workflow Protocol)
The main objective of ultrasound workflow automation (iWorks) is to speed up exam times and
reduce the excessive number of user interface manual key strokes that can lead to repetitive strain
injuries over time. It automates a clinical workflow in common exam protocols in a logical “step by
step” manner. It also prevents missing important parts of examinations as well as decreasing exam
times.
A Protocol Event contains series workflow events (annotation comments, body marks and
measurements) and image modal commands to assist the user in routine ultrasound examinations.
TIP:
•
The system provides different protocol events based on the different application regions.
•
iWorks is an option.
Perform the following procedure:
1.
Input the patient information.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...