10 - 2
Operator’s Manual
10 Physiological Unit Signal
b.
Connect the ECG cable to the ECG module.
c.
Turn on the power supply of the system.
d.
Place the ECG electrodes on the patient’s body (as shown in the following figure).
2.
Tap [Physio] or press the user-defined key for “Physio” to enter physio operation interface.
3.
Switch the imaging modes and display formats, adjusting the parameters to get an optimized
image.
4.
Parameter adjusting:
Tap [ECG] on the touch screen to enable or disable ECG waveform curve. Adjust the [Speed],
[ECG Gain], [Position] and [Invert].
5.
Trigger:
Select the trigger mode, or tap [Real & Trig] to set the trigger time, triggering delay time and
image display format.
6.
Freeze the triggering image and the curve, and then review them.
7.
Tap [Physio] or press the user-defined key for “Physio” to exit ECG mode, and remove ECG
electrodes from the patient.
ECG Triggering
ECG triggering means that image scanning is activated at some time points of ECG signals, thus
obtaining B images at these time points. The triggering image should be in 2D-mode.
When ECG triggering occurs, some marks (frame triggering mark) appear on the ECG waveform
(relative R wave, the time for delay set), indicating the time points when the 2D images are
captured.
TIP:
•
The triggering mark is displayed in both freeze mode and live mode
•
The marks in Dual trigger are in different colors.
•
Triggering function is unavailable if the ECG trace is disappeared. Only the live 2D image can
be triggered.
•
No delay time or time interval shall be less than the time required to scan a single image.
•
If the delay time is longer than a heart cycle, then the heart cycle in the delay time is omitted,
that is to say no trigger is occurred when R waveform is detected in the duration.
I
II
III
N
F
L
R
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...