Operator’s Manual
1 Safety Precautions
WARNING
•
Do not connect the three-wire cable of the system with a two-wire plug
without protective grounding; otherwise, electric shock may result.
•
Do connect the power plug of this system to wall receptacles that meet the
ratings indicated on the rating nameplate. If adapters or multi-functional
receptacles are used, it may cause the leakage current to exceed the
safety requirement.
•
In the environment that patient is 1.5 meters around, connect peripherals to
the auxiliary power outlet, or power the peripherals by auxiliary output
cable or isolation transformer complied with IEC60601-1 or the power input
of the same safety level.
•
DO NOT use power supply of different phases to power peripherals, like
power supply of air-conditioning.
•
When using peripherals not powered by the auxiliary output of the
ultrasound system, or using peripherals other than permitted by Mindray,
make sure the overall leakage current of peripherals and the ultrasound
system meets the requirement of the local medical device electrical
regulation (like enclosure leakage current should be no more than 500 uA
of IEC60601-1), and the responsibility is held by the user.
•
Connect the grounding conductor before turning ON the system.
Disconnect the grounding cable after turning OFF the system. Otherwise,
electric shock may result.
•
For the connection of power and grounding, follow the appropriate
procedures described in this operator's manual. Otherwise, there is risk of
electric shock. Do not connect the grounding cable to a gas pipe or water
pipe; otherwise, improper grounding may result or a gas explosion may
occur.
•
Before cleaning the system, disconnect the power cord from the outlet.
System failure and electric shock may result.
•
This system is not water-proof designed. Do Not use this system in any
place where water or any liquid leakage may occur. If any water is sprayed
on or into the system, electric shock may result or the system may be
damaged. If water is accidentally sprayed on or into the system, contact
Mindray Customer Service Department or sales representative.
•
DO NOT use a probe that has a damaged, scratched surface, or exposed
wiring of any kind. Immediately stop using the probe and contact Mindray
Customer Service Department or sales representative. There is risk of
electric shock if using a damaged or scratched transducer.
•
Do not allow the patient to contact the live parts of the ultrasound system or
other devices, e.g. signal I/O ports. Electric shock may occur.
•
Do not use an aftermarket probe other than those specified by Mindray.
The probes may damage the system causing a profound failure, e.g. a fire
in the worst case.
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...