Operator’s Manual
13 - 1
13
Measurement, Comments and
Body Mark
13.1 Measurement
There are general measurements and application measurement. You can perform measurements on
a zoomed image, cine reviewing image, real-time image, or frozen image. For measurement details,
please refer to the
Advanced Volume
.
WARNING
•
Be sure to measure areas of interest from the most optimal image plane to
avoid misdiagnosis from inaccurate measurement values.
•
To obtain accurate Doppler flow measurement values, make sure the
transmitting beam is not perpendicular to the flow, otherwise false readings
and potential misdiagnosis may result.
CAUTION
•
If an image is unfrozen or the mode is changed during a measurement, the
calipers and measurement data will be cleared from the screen, but the
measurement data will be stored in the report.
•
If the system is turned off or [End] is selected during a measurement, the
data not saved will be lost.
•
In Dual-B imaging mode, the measurement results of the merged image
may be inaccurate. Therefore, the results are provided for reference only,
not for confirming a diagnosis.
Measurement Accuracy
Table 13-1 Error of 2D Images
Parameter
Value Range
Error
Distance
Full screen
Within ±3%
Area Full
screen
Within
±7%
Circ
Full screen
Within ±10%
Angle
Full screen
Within ±3%
Volume Full
screen
Within
±10%
Summary of Contents for Anesus ME7T
Page 2: ......
Page 58: ...This page intentionally left blank ...
Page 154: ...This page intentionally left blank ...
Page 164: ...This page intentionally left blank ...
Page 182: ...This page intentionally left blank ...
Page 190: ...This page intentionally left blank ...
Page 208: ...This page intentionally left blank ...
Page 254: ...This page intentionally left blank ...
Page 264: ...This page intentionally left blank ...
Page 280: ...This page intentionally left blank ...
Page 311: ......
Page 312: ...P N 046 018839 00 5 0 ...