4-21
The following explains the
Calibrator
and
Parameter
.
Parameter Description
No.
Calibrator No. The range is S1-S8.
Std. Conc.
Set point of the calibrator.
Response
Response of the calibrator.
Parameter
Parameters in the calibration formula.
Value
Value of the parameters in the calibration formula.
The following introduces the buttons on the screen.
Buttons
Function
Edit
Edit the parameters in the linear calibration equation.
Enter new values in the
K
and
R
0
text box; click
OK
to save and enter
the
Calibration
screen. The calibration curve and information will be
refreshed in the meantime. Click
Cancel
to cancel the setting.
NOTE
The formula can only be edited when the calibration
rule is linear calibration (single-point, two-point and
multi-point).