4-34
Figure 4-21 Print Features
The following table explains the parameters on the
Print Features
screen.
Parameter Description
Internal
Printer
Select internal thermal recorder to print test result.
External
Printer
Select external printer to print test result. The Appendix lists the
printers the system supports.
Start
internal
online PRN
Print the test result in real time.
The following table introduces the button on the screen.
Buttons
Function
Top
Move the selected test to the top of the list.
Up
Move the selected test upward by one line.
Down
Move the selected test downward by one line.
Bottom
Move the selected test to the bottom of the list.
Cancel
Restore the original printing order.
Return
Click this button to return to the main screen.
4.7.4 Carryover
In
Carryover
screen, you can set some tests between which carryover may exists to
notify operators to rinse the tubes, thus the effect brought about by carryover can be
minimized.
Summary of Contents for BA-88A
Page 1: ...BA 88A Semi auto Chemistry Analyzer Operation Manual...
Page 2: ......
Page 6: ......
Page 18: ......
Page 21: ...III Appendix C External Printer C 1 Appendix D Supplies D 1...
Page 22: ......
Page 28: ......
Page 78: ......
Page 96: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 107: ...P N BA89 20 87038 1 0...