Operating Your Analyzer
6-17
Figure 6-11 “Re-count” screen
Follow the previously introduced procedure to re-analyze the sample in question. The new
result will overwrite the old result while the sample information keeps unchanged.
Automatic saving of analysis results
This analyzer automatically saves a maximum of 10,000 sample results. When the maximum
number has been reached, the newest result will overwrite the oldest.
Parameter flags
If the analysis result is followed by an ”
H
” or “
L
”
,
it means the analysis result has
exceeded the upper or lower limit of the reference range.
If you see
***
as opposed to the result, it means the result is either unreliable or out of the
operating range.
If the WBC result is less than 0.5
×
10
9
/L, this analyzer will not perform the differential
analysis and all the related parameter values will be non-numeric (
***
)
.
z
The result of the background check will not be flagged.
Histogram flags
The system will flag abnormal histograms.
Abnormal WBC histograms will be flagged by one of the markings: R1, R2, R3, R4 and
R
m.
R
1
: indicates abnormality on the left side of the lymphocyte hump and possible presence of
Summary of Contents for BC-2800
Page 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Page 2: ......
Page 12: ......
Page 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Page 62: ......
Page 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Page 92: ......
Page 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Page 196: ......
Page 248: ......
Page 252: ......
Page 266: ......
Page 284: ...Appendices E 10 Hemoglobin Concentration ...
Page 286: ...P N 2800 20 28795 2 0 ...