Using the Calibration Programs
9-10
Verifying new calibration factors
Press [MENU] to exit the “
Calibrator
” screen, a message box will pop up to confirm the new
calibration factors, as Figure 9-9 shows.
Figure 9-9 A message box to confirm the new calibration factors
CLICK
”
Yes
” to save the new calibration factors and enter the “
Count
” screen from the
system menu.
At the “
Count
” screen, run the calibrator or a normal-level control at least 3 consecutive times
and calculate the mean of the results. Compare the obtained means to the expected means.
Other operations
Printing new calibration factors
Press [PRINT] to print out the current calibration factors.
Exiting the ”
Calibrator
” screen
Press [MENU] to exit to system menu.
9.3.3 Fresh Blood Calibration
ENTER
the administrator password in the “
Password
” screen. Press [MENU] to enter the
system menu.
Figure 9-10 System menu