Appendices
E-4
and V
Message description unit
Field
name
Coding Description
CTR
Data package of normal analysis
results
QCR
Data package of QC running
MD
QCC
Data package of QC standard
Programming
The BC-2800 must communicate with the external computer when the Handshake is on in
following procedures:
1. During the communication connection, the BC-2800 sends an ENQ (10 Hex), then waits
up to 4 seconds for the external computer to respond. If the external computer does not
respond, then one more ENQ (10 Hex) is tried. If it fails again, the analyzer aborts the
transmission and reports a transmission error;
2. The external computer must respond by sending an ACK (06 Hex);
3. After the ACK (06 Hex) is received, the analyzer then sends data block;
4. After the sending, the BC-2800 sends an ETX (0F Hex), then waits 4 seconds for the
external computer to respond. If no response is received, one more ETX (0F Hex) is sent.
The BC-2800 waits 4 seconds before giving up and gives alarm of communication error.
E.3.2 Sample Data format
FD
field
FD definition of analysis parameter
FD Description
Sample information
SampleID Sample
ID
Mode
Analysis Mode (0:whole blood-all parameter; 1: predilute-all parameter;
2: whole blood- WBC/HGB; 3: predilute - WBC/HGB; 4: whole
blood-RBC/PLT; 5: predilute - RBC/PLT)
TestTime
Test time, format: YYYY-MM-DD HH: MM: SS (if the digit is less than
10, add a 0 at the ten’s place)
Name Name
Summary of Contents for BC-2800
Page 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Page 2: ......
Page 12: ......
Page 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Page 62: ......
Page 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Page 92: ......
Page 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Page 196: ......
Page 248: ......
Page 252: ......
Page 266: ......
Page 284: ...Appendices E 10 Hemoglobin Concentration ...
Page 286: ...P N 2800 20 28795 2 0 ...