Installing Your Analyzer
4-3
4.2.3 General Environment
Operating temperature: 15
℃
to 30
℃
.
Relative humidity: 30% to 85%.
Atmospheric pressure: 60 kPa to 106 kPa.
The environment should be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference.
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close.
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
z
Do not place the analyzer in a flammable or explosive environment.
z
Do not place any container on the top of the analyzer.
z
If the ambient temperature is outside the specified operating range, the
analyzer will alarm you for abnormal ambient temperature and the analysis
results may be unreliable. See Chapter 11 Troubleshooting Your Analyzer for
solutions.