Customizing the Analyzer Software
5-14
To increase (or decrease) the gain
1. Enter the administrator password as introduced in
Chapter 5.2.1
.
2. At the ”
Gain
” screen and
ENTER
the desired gain into the “
WBC (WB)
”
,
as
Figure 5-24
shows, or “
WBC (PB)
”, as Figure 5-25 shows.
Figure 5-24 Setting WBC (WB) gain
Figure 5-25 Setting WBC (PB) gain
Setting the RBC gain
When the difference between the actual MCV result and the expected result exceeds 6%,
you need to change the RBC gain.