10-2
To clean your equipment, follow these rules:
1. Turn off the pump and disconnect the AC power source line.
2. Wipe the display screen after soft cotton balls absorb an appropriate amount of
detergent.
3. Use a piece of soft cloth which absorbs a modest amount of cleaning agent to
wipe the surface of the device.
4. When necessary, use a piece of cloth to wipe off any excess cleaning agents.
5. Place the pump in a cool and ventilated environment to dry.
10.3 Disinfection
The operation of disinfection may cause certain damage to the syringe pump. You
are recommended to disinfect only when it is necessary in your desired
maintenance plan. Clean the equipment before disinfection.
The recommended disinfectants include: Ethanol (70%), Isopropanol (70%),
glutaraldehyde-type 2% liquid disinfectant.
CAUTION
Never use EtO or formaldehyde for disinfection.
Do not conduct high pressure or high temperature disinfection for the
syringe pump and its accessories.