4 - 4
4.8
Unloading the Infusion Set
WARNING
•
To prevent free flow, ensure that the roller clamp or Robert clamp is closed
before removing the infusion set from the pump.
•
It is recommended that the infusion sets be changed every 24 hours (or as per
national hygiene regulations or manufacturer’s instructions). If infusion sets
not recommended by this manual are used, adjust the fixing site every four
hours.
•
It is recommended that the disposable enteral nutrition sets be changed
every 24 hours.
•
If the infusion sets are not changed within the recommended time, the
accuracy of the infusion may be affected.
4.9
Turning Off the Pump
CAUTION
•
Press and hold the power switch for no less than 10 seconds to forcibly shut
down the pump if it could not be shut down normally. This may cause loss of
patient data.
NOTE
•
Turning off the pump does not disconnect the pump from the AC mains. To
completely disconnect the power supply, unplug the power cord.