3-12
4. Open the relief valve and vent a certain standard gas or gas mixture. Then adjust the
relief valve and make sure the flowmeter reading is stable and within 10 and 50 L/min.
5. In
the
[
Calibrate AG
] menu, the concentration and flowrate of each measured gas are
displayed.
If the difference between the measured gas concentration and the actual one is t
very small, a calibration is not needed.
If the difference is great, a calibration should be performed. Select [
Calibrate >>
]
to enter the calibrate menu.
6. Enter the vented gas concentration. If you use only one gas for calibration, set other
gases’ concentration to 0.
7. Select
[
Start
] to start calibration.
8. If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed. If the calibration failed, the message [
Calibration Failed!
] is displayed. In
this case, perform another calibration.
CAUTION
z
Calibrate the O
2
module, If it has been transported for long distance.
Summary of Contents for BeneView T5