3-14
Block the gas inlet of the module or watertrap. The sidestream
CO
2
flowrate is slower than 10ml/min and an alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
Yes No
The displayed CO
2
value is within 6
±
0.05
%
.
Yes No
Microstream CO
2
test
Block the gas inlet of the module or watertrap. An alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
Yes No
The displayed CO
2
value is within 6
±
0.05
%
.
Yes No
AG test
When AG flowrate is slower than 10ml/min, an alarm of AG
Airway Occluded is given. It indicates that there is no leakage.
Yes No
The fan inside the AG module works properly.
Yes No
The measurement accuracy of CO
2
, N
2
O, O
2
and AA (AA
represents an anaesthetic agent) meets the product specifications
in the Operator’s Manual.
Yes No
3.3 Power On Test
This test is to verify that the patient monitor can power up correctly. The test is passed if the
patient monitor starts up by following this procedure:
1. Insert two batteries in the battery chamber and connect the patient monitor to the AC
mains, the AC mains LED and battery LED light.
2. Press the power on/off switch to switch on the patient monitor. The operating status
LED lights up, and the technical and physiological alarm lamps light blue and red
respectively.
3. After the start-up screens are displayed, the system sounds a beep indicating the self test
on alarm sounds is passed. At the same time, the alarm lamp turns from yellow to red,
and then turns off together with the technical alarm lamp. This indicates that the self test
on alarm lamps is passed.
4. The patient monitor enters the main screen and start-up is finished.
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......