3-17
3.4.5 Temp Test
Tool required:
Resistance box (with accuracy above 0.1
Ω
)
Follow this procedure to perform the test:
1. Connect the two pins of any Temp connector of a module to the two ends of the
resistance box using 2 wires.
2. Set the resistance box to 1354.9
Ω
(corresponding temperature is 37ºC).
3. Verify that the displayed value is within 37 ± 0.1ºC.
4. Repeat steps 1 to 3 and verify another temperature channel.
3.4.6 IBP Tests
IBP Performance Test
Tool required:
Medsim300B patient simulator, MPS450, or other equivalent device
Dedicated IBP adapter cable for test (P/N 009-002199-00 for Medsim 300B, P/N
009-002198-00 for MPS450)
Follow this procedure to perform the test:
1. Connect the patient simulator with the pressure module.
2. Make the patient simulator outputs 0 to an IBP channel.
3. Press the Zero Key on the module to make a zero calibration.
4. Configure the patient simulator as P (static) = 200 mmHg.
5. The displayed value should be within 200 ± 4 mmHg.
6. If the error is beyond ±4 mmHg, calibrate the pressure module. If the IBP module was
calibrated with a dedicated reusable IBP sensor, check the calibration together with this
IBP sensor.
7. Make the patient simulator output 120/80 mmHg ART signal and 120/0 mmHg LV
signal to the IBP channel and check that the IBP wave is displayed correctly.
8
Repeat the steps above for all the IBP channels.
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......