4-8
4.6.7 Output Interface Failures
Symptoms
Possible Cause
Corrective Action
Respective output
disabled
1. Select [
Main Menu
]
→
[
Analog Output
Setup
]
→
set [
Analog Output
] to [
On
].
USB_Hub board cable
loose
1. Check that cable between USB_Hub
board and mother board is properly
connected.
2. Check that connecting cables and
connectors are not damaged.
USB_Hub board failure
Replace the USB_Hub board.
No analog signals or
nurse call signals are
issued
Mother board failure
Replace the mother board.
Cable defective or
poorly connected
1. Check that cable between USB_Hub
board and mother board is properly
connected.
2. Check that connecting cables and
connectors are not damaged.
USB_Hub board failure
Replace the USB_Hub board.
Device with USB port
does not function
(Assume that the
peripheral devices are
good)
Mother board failure
Replace the mother board.
4.6.8 CF Card Problems
Symptoms
Possible Cause
Corrective Action
Wrong CF card or small
memory space
Use only SanDisk-manufactured CF storage
cards. Those with 1GB memory space are
recommended.
CF card full; data error;
CF card error
Format CF card on PC.
CF card failure
Replace the CF card.
Cable defective or
poorly connected
1. Check that the cable between CF card
board and mother board is correctly
connected.
2. Check that connecting cables and
connectors not damaged.
CF card board failure
Replace the CF card board.
CF card malfunctions
Mother board failure
Replace the mother board.
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......