System Structure and Assembly/Disassembly
4-49
3)
Remove 2 M3X8 countersunk screws for each transducer mounting bracket
and shielding plate, remove 2 M2.5X8 panhead screws for the transducer
socket and transducer mounting bracket, and remove the transducer mounting
brackets (6 in total).
4)
Remove 12 M3x8 screws for the transducer board and shielding plates, and
separate the shielding plates and transducer board.
4.3.20 Replace the inlet dust net
a)
Loosen two nondetachable screws for two inlet dust nets and main unit housing,
and take out the inlet dust net from the left side of the system.
M3X8
countersunk
screw
M2.5X8 panhead
screw
Transducer
mounting
bracket
Transducer board
Shielding plate