Image Optimization 5-41
Image parameters sett
ing, for details, please refer to “5.11.3.1 Image Parameters ”.
Image zooming, for details, please refer to “5.11.3.2 Image Zooming”.
Image rotation, for details, please refer to “5.11.3.3 Rotating the Image”.
Measurement, comment, and body mark, for
details, please refer to “5.11.3.4
Measurement, Comment, and Body Mark
5.11.3.1 Image Parameters Adjusting
In image viewing mode, you can adjust the following parameters:
Image size
Click [Actual Size] to display the image in its actual size.
Click [Fit Size] to display the image in an appropriate size according to the current
window. Also this is the system default format.
Tint map
Click the [Tint Map] on the menu to adjust it.
Ruler
Click [Ruler] on the menu to hide or show the ruler around the image.
5.11.3.2 Image Zooming
Press <Depth/Zoom> knob on the control panel to enter image zooming mode. Rotate the
knob to zoom in/out the panoramic image.
Rotate clockwise to zoom in the image.
Rotate anticlockwise to zoom out the image.
Roll the trackball to change position of the magnified image.
Press the <Depth/Zoom> knob again to exit the zoom mode.
When image displayed is bigger than the image area, the thumbnail is automatically
displayed.
5.11.3.3 Rotating the Image
For the convenience of viewing the image, you can rotate the image by clicking [Rotation] on
the image menu.
5.11.3.4 Measurement, Comment, and Body Mark
In iScape image viewing status, you can perform measurement, comment, and body mark.
The operations are the same as that of B mode.
NOTE:
1. The measurement accuracy for the spliced image may be degraded, exercise
caution when measurements are performed on an iScape image.
2.
If there is a trace during the retracing, please don’t perform measurement
across the trace.
5.11.3.5 Evaluate Image Quality
Many variables may affect the overall image quality. It is important to evaluate the image
content and quality before an image is used for diagnosis or measurements.
NOTE:
iScape panoramic imaging is intended for well-trained ultrasound operators or
physicians. The operator must recognize image artifacts that will produce a sub-
optimal or unreliable image.
The following artifacts may produce a sub-optimal image. If the image quality cannot meet
the following criteria, you shall delete the image and do image acquisition again.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...