12-34 Probes and Biopsy
Figure 1 Figure 2
To enter/exit iNeedle
To enter iNeedle
Click [iNeedle] item in B page on the screen.
Or, you can assign a user-defined key for entering iNeedle.
Open iNeedle in Biopsy status
1. Perform scanning and locate the target, press <Biopsy> to enter the screen.
2. Click [iNeedle] to enter the status; available adjusting parameters are displayed on the
menu.
To exit iNeedle
Press the user-defined key or click [iNeedle] to exit the status and enter B mode.
Needle Steer
Description
This function adjusts the biopsy needle angle via changing the steer angle of
the scan line. The iNeedle affecting region changes correspondingly.
Operation
Click [Needle Steer] item on the screen.
B/iNeedle
Description
This function is used to display B image and iNeedle image synchronously.
Operation
To turn on or off the function, click [B/iNeedle] on the screen.
Tips: iZoom (full-screen magnification) is available on iNeedle status.
12.2.8 Removing the Needle-guided Bracket
NGB-001/ NGB-002/ NGB-003/ NGB-005 metal/needle un-detachable needle-guided
bracket:
While holding the transducer and the needle-guided bracket, open the Grip knob of the
needle-guided bracket.
NGB-001 Metal-needle detachable
1.
Pull the lock pin and open up the V-shaped cover to expose the needle.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...