12-36 Probes and Biopsy
NGB-005
While holding the transducer and the needle-guided bracket, open the Grip knob of the
needle-guided bracket.
NGB-007
Metal
1) Screw the nut of the guiding block and remove the guiding block slightly along
the direction of the needle’s tail.
2) Separate the residual part of the needle-guide bracket and the transducer from
the needle.
3) Screw the pinch nut of the bracket, and remove the needle-guided bracket
from the transducer.
Plastic
1)
Remove the guiding block slightly along the direction of the needle’s tail.
2) Separate the residual part of the needle-guide bracket and the transducer from
the needle.
3) Remove the support of needle-guided bracket from the transducer.
NGB-009
(1) Turn on the knob of fixing the needle and separate the needle from the needle-
guided bracket.