Acoustic Output Reporting Table (60601-2-37) C-87
Transducer Model: 75L50EAV
Imaging Mode: B-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.97
0.13
0.31
0.14
Index component value
0.13
0.13
0.31
0.13
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
2.44
P
(mW)
22.00
22.00
10.00
P
1×1
(mW)
4.58
4.58
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.17
z
pii
,
α
(cm)
1.17
f
awf
(MHz)
6.31
5.99
5.95
5.16
Other
Informatio
n
prr
(Hz)
5,086.0
0
srr
(Hz)
41.00
n
pps
1.33
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
131.39
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
3.44
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
5.72
p
r
at
z
pii
(MPa)
3.15
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
Focus position
1.5cm
10.0cm
10.0cm
10.0cm
Working Frequency
7.5
5
5
H10.0
PRF
5086
5086
5086
5086
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...