Image Optimization 5-21
Effects
Generally, increasing the acoustic power will increase the brightness and
contrast of the image as well as the force of penetration.
Impacts
You should perform exams according to actual situation and follow the ALARA
Principle.
Baseline
Description
Refers to the area where the velocity is zero in the spectrum.
Operations
Press the <Baseline> key on the control panel and rotate the Multifunctional
Knob on the left to adjust. The adjustment range is -4 to 4.
Effects
To change the range of flow velocity to optimize the image.
PW Steer
Description
This function provides adjustment on angles for sampling line.
Operations
Adjust through the [Steer] item on image menu.
Effects
This feature is used to steer the direction of the beam so as to change the
angle between the beam and flow direction with immobility of the linear probe.
Values of steer angles vary by probe.
Steer is available only for linear probes
Invert
Description
This function is used to set the display manner of spectrum.
Operations
Turn on or off the function through the [Invert] item on the image menu.
Scale
Description
This function is used to adjust the speed range of flow, which is adjusted
through PRF in the system.
The real-time PRF value is displayed in the image parameter area in the
upper right corner of the screen.
Operations
Press the <Scale> key on the control panel and rotate the Multifunctional
Knob on the left to adjust.
Effects
To provide a much clearer color flow image.
Use low PRF to observe low-velocity flows, and use high PRF to observe
high-velocity flows.
Impacts
Aliasing may occur if low velocity scale is used and high velocities are
encountered.
Low velocities may not be identified when a high velocity scale is used.
Wall Filter
Description
It filters out low-velocity signals to provide effective information, and this
function is used to adjust the filtered frequency. The real-time value is
displayed in the image parameter area in the upper right corner of the screen.
Operations
Select through the [WF] item on the image menu.
There are 0-6 levels of wall filter function provided.
Impacts
Signals of low-velocity flow may be missing.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...