Image Optimization 5-1
5
Image Optimization
WARNING:
1. The images displayed in this system are only reference for
diagnosis. Mindray is not responsible for the correctness of
diagnostic results. It is the responsibility of the clinician,
who performs the exam, to capture the correct diagnostic
results.
2.
In Dual-B imaging mode, the measurement results of the
merged image may be inaccurate. Therefore, the results are
provided for reference only, not for confirming a diagnosis.
5.1
Switching Between Image Modes
Switch the image modes. For details, refer to 2.6.3 Control Panel.
5.2
Basic Operations
Before optimizing the image by adjusting image parameters, adjust the brightness and
contrast of the display to the best.
Requirement
Available Operations
To modify the
brightness
Adjust [Gain].
Adjust TGC
Adjust [Acoustic Power] (try adjusting gain before
increasing the acoustic power)
To modify gray scale
image effect
Adjust [Dynamic Range]
Adjust [Map]
Adjust [Persistence]
Adjust [iClear]
To increase frame
rate
Decrease [Depth]
Decrease the [Focus Number]
Decrease the [FOV]
Decrease [Line Density]
Turn on [High FR] in harmonic mode
Adjusting through Image Menu:
Press <Menu> or multifunctional knob to open the menu; use the trackball and <Set> or
the multifunctional knob to adjust.
Adjusting through control panel:
Trackball, control panel key, knob or sliders.
Summary of Contents for DP-50 Exp Vet
Page 2: ......
Page 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Page 42: ......
Page 68: ......
Page 128: ......
Page 148: ......
Page 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Page 180: ......
Page 220: ......
Page 224: ......
Page 236: ......
Page 242: ......
Page 248: ......
Page 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Page 343: ...P N 046 017713 02 1 0...