Installation Menu
Repair Information
3 - 54
The following table describes the Installation Menu structure:
MENU TITLE ON
SCREEN
MENU CHOICES
DEFAULT
ACTIONS/
COMMENTS
Save Current
Select to save current
settings as defaults.
Select Language
English, German,
French, Italian,
Spanish, Portuguese,
Dutch
English
Select to change
language.
Date Format
M/D/Y, D/M/Y, Y/
M/D
M/D/Y
Select to change date
format.
Time Format
12 hour, 24 hour
12 hour
Select to change time
format.
NIBP Timeout
15min, 30min,
45min, 60min
15min
Selects time old and
data is removed from
screen.
Temperature Units
ºC, ºF
ºF
Select to change
temperature units.
Weight Units
kg, lbs
lbs
Select to change
weight units.
Height Units
cm, Ft/inches
inches
Select to change
height units.
CO
2
Units
mmHg, %, kPa
mmHg
Select to change CO
2
units.
ECG Standard
AHA, IEC
AHA
Select to change ECG
lead standard.
Notch Filter
50Hz, 60Hz
60Hz
Select to change notch
filter frequency.
ESU Filter
Auto, Disable
Auto
Select to change ESU
Filter setting.
SpO
2
Sensor Off
Audio
Off, Once, Repeat
Off
Select to change
SpO
2
Sensor Off
Audio.
SpO
2
Tones
Standard Tones,
Alternative Tones
Standard Tones
Select to change the
SpO
2
tones.
Enable Arrhythmia All
Off Selection
Yes, No
No
Select to enable or
disable the Arrhythmia
All Off menu
selection.
Apnea Latch
On, Off
On
Select to turn apnea
alarm latching on or
off.
Enable Silence All
Permanent Selection
Yes, No
No
Select to enable or
disable the Permanent
Audio Off menu
selection.
Auxiliary Output
Nurse Call, Analog
Output, Defib Sync
Nurse Call
Select to change the
Auxiliary Output.
Nurse Call Mode
Normally Open,
Normally Close
Normally Open
Select to change
Nurse Call Mode.
Set up Serial Port 1
None, DIAP, Gas
Module
None
Select to set up a
serial output protocol
port.